Chapter 14: Infant and Pediatric Devices

Mosbys Respiratory Care Equipment 9th Edition

$2.99

Chapter 14: Infant and Pediatric Devices

 

Complete Chapter Questions With Answers

 

Sample Questions Are Posted Below

 

MULTIPLE CHOICE

 

  1. What is the respiratory rate available on the CareFusion SiPAP?
a. 1 to 100 breaths/minute c. 1 to 150 breaths/minute
b. 1 to 125 breaths/minute d. 1 to 200 breaths/minute

 

 

ANS:  C

The respiratory rate on the CareFusion SiPAP is adjustable from 0 to 150 breaths per minute.

 

PTS:   1                    REF:   Page 466, Table 14-1

 

  1. When is the oxygen blender on the CareFusion Infant Flow Nasal CPAP System recalibrated?
a. When the calibration control is activated
b. When the ventilator is turned on
c. When there is a change in the delivered fractional inspired oxygen
d. At start-up and every 12 hours

 

 

ANS:  B

The oxygen blender is automatically recalibrated on start-up.

 

PTS:   1                    REF:   Page 465

 

  1. Which of the following are power sources for the Cardinal Health Infant Flow Nasal CPAP System?
  2. 12-V direct current battery
  3. 20-psig compressed-air source
  4. 50-psig oxygen source
  5. Alternating current electrical source
a. 1 and 4 c. 2, 3, and 4
b. 1 and 2 d. 1, 2, 3, and 4

 

 

ANS:  D

The continuous positive airway pressure system uses two high-pressure gas sources (50 psig each), an alternating current electrical source, and a 12-V direct current battery as a back-up electrical power source.

 

PTS:   1                    REF:   Page 464

 

  1. The Cardinal Health Infant Flow Nasal CPAP System driver can operate on battery power for _____ hours.
a. 2 c. 6
b. 4 d. 10

 

 

ANS:  B

A lead acid battery is contained within the driver. With a full charge, the driver unit operates for 4 hours on battery power.

 

PTS:   1                    REF:   Page 464

 

  1. What is the maximum amount of time an alarm condition must persist on the CareFusion Infant Flow Nasal CPAP System for an alarm to be delayed when the ventilator is being set up?
a. 30 seconds c. 90 seconds
b. 60 seconds d. 120 seconds

 

 

ANS:  D

Alarms are delayed for 120 seconds after entering the main screen to allow time for ventilator set up.

 

PTS:   1                    REF:   Page 465