Chapter 15: Transport, Home Care, and Noninvasive Ventilatory Devices

Mosbys Respiratory Care Equipment 9th Edition

$2.99

Chapter 15: Transport, Home Care, and Noninvasive Ventilatory Devices

 

Complete Chapter Questions With Answers

 

Sample Questions Are Posted Below

 

MULTIPLE CHOICE

 

  1. The pneumatic control logic of the Airon pNeuton requires a flow of _____ L/min to operate.
a. 1 c. 3
b. 2 d. 4

 

 

ANS:  C

The internal control logic requires a flow of 3 L/min. The additional 1 L/min on the Airon pNeuton model A is to power the patient-disconnect alarm.

 

PTS:   1                    REF:   Page 515

 

  1. What alarm(s) is (are) on the Airon pNeuton ventilator?
  2. Low positive end-expiratory pressure/continuous positive airway pressure
  3. Patient disconnect
  4. High patient pressure
  5. Low oxygen inlet pressure
a. 2 only c. 2 and 4
b. 1 and 4 d. 1, 2, 3, and 4

 

 

ANS:  C

The pNeuton has two functional alarms that are always active: a low oxygen inlet pressure alarm that activates at pressures less than 30 psi and a patient-disconnect alarm that sounds when a disconnect situation is sensed by the ventilator.

 

PTS:   1                    REF:   Page 516

 

  1. Which of the following modes of operation are offered by the Airon pNeuton?
  2. Volume-controlled intermittent mandatory ventilation
  3. Pressure-controlled intermittent mandatory ventilation
  4. Continuous positive airway pressure
  5. Pressure-regulated volume control
a. 1, 2, and 3 c. 1, 3, and 4
b. 2, 3, and 4 d. 2 and 3

 

 

ANS:  A

The pNeuton offers nonsynchronized volume-controlled intermittent mandatory ventilation or pressure-controlled intermittent mandatory ventilation and positive end-expiratory pressure/continuous positive airway pressure.

 

PTS:   1                    REF:   Pages 516-517

 

  1. The bias flow on the Airon pNeuton is _____ L/min.
a. 3 c. 8
b. 5 d. 10

 

 

ANS:  D

The pneumatic system of the pNeuton provides a continuous flow of gas through the circuit of approximately 10 L/min to meet the inspiratory demand during spontaneous breaths.

 

PTS:   1                    REF:   Page 517

 

  1. The internal direct current battery of the Bio-Med Crossvent operates for approximately ____ hours.
a. 2 c. 6
b. 4 d. 8

 

 

ANS:  C

The internal direct current battery is operational for approximately 6 hours when fully charged.

 

PTS:   1                    REF:   Page 517